FRP Series/Full Range Programmable Humidity Chambers
Control System
The FRP series is supplied with a Pro-600 Programming Control System
as standard. Time vs. temperature and relative humidity profiles are
programmable through the Pro600. It has two sets of controls featuring
proportional, integral and derivative parameters, plus reset windup inhibit
and reset clipping facilities. This control system provides control setting
and indication resolution of 0.1° and +/- 0.1% relative humidity.
This controller has 200 program steps for up to 99 recipes and 16 programmable
events. An RS-232 communications port is included along with a menu driven
software package for the customer's host computer. A host computer can
communicate with the Pro-600 programmer via EIA RS-232C serial communications
protocol. This interface is provided as standard in the Pro-600 control
system. A free diskette for the Pro600, containing programs for setting
up communications is available upon request. Other interface protocols,
RS-422 and IEEE-488, are optionally available.
Recorder
Blue M's B M 7000 microprocessor- based 10" diameter circular chart
recorder is standard. It is capable of measuring, displaying, and recording
dry bulb temperature and % of relative humidity. All recorder functions
are easily configured via the keys on the front panel. A special swing-out
mounting bracket allows the recorder/controller to be pivoted from the
side of the control compartment to the front if desired.
Humidity Control
Method
Rather than using a moisture injection system, Blue M temperature/humidity
chambers use a true vapor pressure system for humidity control. Chamber
air is humidified or de-humidified using an integral water reservoir.
By controlling the temperature of the water in the reservoir (thereby
changing the vapor pressure above the liquid) moisture is added or subtracted
from the chamber air. This technique closely approximates natural humidity
control. Superior accuracy and stability, out-stripping any moisture
injection method, are achieved using this method.
Horizontal Air Circulation
A high efficiency air circulation system, delivering horizontal airflow,
is an integral part of chamber construction. This direct drive design
has an extended shaft electric motor coupled to a stainless steel centrifugal
blower wheel. Smaller models feature single blowers; the larger units
have dual blowers.
Solid State Water Levelers
Chamber water usage is regulated by solid state water level controls
of Blue M design and manufacture. Leveler control circuitry is designed
to automatically control the Blue M water purification option. An integral
reducing pressure regulator is provided for reliable operation on high
pressure, in house de-ionized (D.I.) water systems Automatic Drain System
The chamber water reservoir is protected by a temperature sensitive switch
that automatically drains the water reservoir when the chamber temperature
exceeds the boiling temperature of water.
Temperature/ Humidity Measurement
All of the models incorporate a relative humidity transmitter, or dry
sensor, to measure chamber relative humidity. These sensors offer excellent
control and display accuracy while eliminating the routine maintenance
involved in wet/dry bulb sensing method.
The temperature measured using a separate 100 Ohm platinum Resistant
Temperature Detector (RTD).
Heating Two separate heating systems are supplied in Blue M temperature/humidity
chambers; one is a dry bulb (air temperature) heater and the other is
a wet bulb (humidity) heater. The dry bulb heater is a Blue M "Eterna" heating
element. It has an open wire resistance heating element which provides
fast, accurate response to controller heating demands. Stainless steel
sheathed heating elements are used for wet bulb heating -- immersed in
the chamber water reservoir. Electric power is regulated by a solid state
relay and controlled from the programmer/temperature controller.
All Stainless
Steel Chamber Construction
Blue M temperature/humidity chambers look good in any operating environment
and they are built to last -- chamber exteriors are fabricated of Type
304 stainless steel with a brushed finish; interior surfaces are fabricated
of 316 stainless steel. All seams are heli-arc welded and passivated
for strength and durability. Gasketed service access panels insure vapor
barrier integrity and maximum insulation effectiveness. A nominal insulation
thickness of four inches is used to minimize heat gain/loss across the
insulated wall. Type 316 stainless steel is used for interior construction
to provide excellent corrosion resistance and long chamber life. Like
the exterior, all interior seams are heli-arc welded and passivated to
insure vapor barrier integrity and prevent insulation damage.
Door Construction
All chambers include inner and outer chamber doors as standard. The
inner glass door provides the primary seal for the temperature/humidity
environment and permits the work in the chamber to be observed during
operation. Inner and outer doors are sealed around their entire perimeter
by silicone sponge rubber gaskets. The chamber inner door is supplied
with a wiper blade to clear window condensation for viewing.
Refrigeration
System
A dual capacity cooling system -- a noncycling hermetic type -- utilizing
a non CFC refrigerant, fulfills refrigeration requirements. A service
panel is supplied in the left hand side of the chamber's machinery base.
The use of desuperheater heat exchangers make it possible to dissipate
live load wattage and/or lower the chamber temperature.
Test-Cable Lead-in
Port
A 3" diameter port with rubber stopper is included as standard
for passing cabling, etc, into the chamber workspace for testing purposes.
Shelves
Two shelves constructed of 304 electro-polished stainless steel are
included as standard with each temperature/humidity chamber.
Chamber Options
Several optional accessories are available to help tailor Blue M temperature/humidity
chambers to your application requirements.
Water Purification Unit
The WP-4800 water purification unit provides 48 gallons of conditioned
water for use in the chamber's humidity and wet bulb measurement systems.
The major components of the unit are a Type 304 stainless steel water
tank, a quiet electric water pump, a sediment filter and a demineralizing
cartridge. Provision for the WP-4800 is designed into the chamber control
circuitry -- the pump is activated by water leveler demand. The chamber
reservoir may be drained into the tank of the WP when the chamber needs
to be drained. This water is passed through the sediment and demineralizing
cartridges before being reused to remove contaminants. This feature aids
the chamber in avoiding corrosion damage from the accumulated contaminants.
Truline
Circular Chart Recorder
Certain testing criteria require greater chart resolution than obtainable
using Blue M's standard circular chart recorder. For these applications,
a Honeywell "TRULINE" circular chart recorder option is available.
This option has a 12" diameter chart that provides 34% greater chart
area than a standard 10" diameter recorder chart. The unit prints
its own chart with a thermal dot matrix stylus at the same instant that
it is printing the process temperature and relative humidity.
Caster/Leveling
Legs
Large, phenolic, swivel casters and sturdy leveling legs can be installed
to enhance the mobility and setup of Blue M chambers. The legs are threaded
rods, with locking nuts and no-mar phenolic feet. Once the unit is in
position, the leveling legs may be adjusted to level the unit and pick
the casters up off the floor.
Drip Shield
Certain high temperature/high humidity conditions can produce condensation
on the ceiling and walls of the chamber workspace. The addition of the
Type 316 stainless steel drip shield prevents condensation on the ceiling
of the chamber from falling on the workload and causing damage (it is
routed down the back wall of the chamber, where it returns to the chamber
water reservoir).
Glove Ports or Port Plugs
Glove ports can be installed in the inner door to allow the operator
to manipulate the product under test. This option includes neoprene rubber
gloves installed through the inner glass door, below the viewing window.
Some processes require small parts or tools to be introduced to the work
space during chamber operation. In these cases, port plugs are installed
through the inner door. These are 6" diameter stainless steel plugs,
held in place with DeStaco clamps. Like the glove ports, these entrances
are installed below the viewing window.
Use of these options is not recommended
at chamber temperatures over +50°C.
Lead-in
Ports
Additional lead-in ports may be installed in the chamber in addition
to the standard port. Larger port diameters are available upon special
request.
Extended Temperature Range
High Temperature: the temperature range of Blue M temperature/humidity
chambers can be extended to +150°C with this option. It permits dry
bulb "temperature only' testing at temperatures above the normal
relative humidity operating range. Low Temperature: the temperature range
of the Blue M temperature/humidity chambers can be extended to -40°C
with this option. It permits dry bulb "temperature only" testing
at temperatures below the normal relative humidity operating range.
The chamber water reservoir is protected by a temperature sensitive
switch that automatically drains the water reservoir when the chamber
temperature exceeds the boiling temperature of water.
Pro-750 Programmer
Controller
All chambers are available with a Blue M PRO-750 Programmer/ Temperature
controller installed in lieu of the standard Pro-600 controls. Please
consult with your local Blue M representative for details.
Computer Interfaces
Blue M programmable controllers come with an with an RS-232 Serial
Digital Communications port as standard. However, some customers find
that other interface protocols are desired For their particular use of
the units. Blue M offers an RS-422/4X5 Serial Communications option and
an IEEE-488 Parallel Communications option for those customers not wishing
to use RS-232.
Engineered-To-Order Options
A wide variety of Engineered-To-Order (ETO) options are available to
configure your chamber for its intended use. We can assist you with your
needs.
Model #
-15 to +93C |
Inside Dimensions Inches (CM) |
Overall Dimensions Inches (CM) |
Full Load Amps |
Voltage 60Hz AC |
Cubic Feet Capacity (Liters) |
| W |
D |
H |
W |
D |
H |
| FRP-09B |
25 (64) |
25 (64) |
25 (64) |
64 (160) |
39 (94) |
71 (180) |
37 |
208V/1PH |
9 (262) |
| FRP-09C |
25 (64) |
25 (64) |
25 (64) |
64 (160) |
39 (94) |
71 (180) |
39 |
240V/1PH |
9 (262) |
| FRP-13B |
37 (94) |
25 (64) |
25 (64) |
751/2 (191) |
39 (94) |
71 (180) |
41 |
208V/1PH |
13 (385) |
| FRP-13C |
37 (94) |
25 (64) |
25 (64) |
751/2 (191) |
39 (94) |
71 (180) |
44 |
240V/1PH |
13 (385) |
|